See II.2.4
This contract concerns a framework agreement for the supply, installation and implementation of laboratory equipment chemistry/immunochemistry including additional peripherals, related reagents, consumables and maintenance.
The analytical system offered will consist of a connected chemistry and immunochemistry analyzer including the middleware associated with the LIS (GLIMS version 9). A physical link between analyzers is preferable to limit the number of links.
In order to ensure that the system is operational 24/7, the tenderer should propose a deduplication of the analytical part (chemistry and immunochemistry).
The analyzers can handle a daily load of at least 3000 chemistry and 500 immunochemistry tests. During a peak period of 2 hours a day, a load of 800 chemistry and 200 immunochemistry tests without loss of TAT is possible. Please explicitly confirm that this is being met. The number of annual tests (guiding only) amounts to 550.00 chemistry and 85,000 immunochemistry tests.
The current laboratory Lab Water Purification System: Milli-Q CLX 7040 (installation year 2021) can be connected to the analyzers. Please explicitly confirm that the quality standard of this Lab Water Purification System can ensure the perfect operation of the analysers offered.