Oslo University Hospital will procure two units and one option (as per operation 2) single door steam sterilisers (approx. two sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.
Sterilisation device/container dimensions: (L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15 mm), (L60 mm x B40 mm x H20 mm), (L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34.5 cm.
Oslo University Hospital will procure two units and one option (as per operation 2) single door steam sterilisers (approx. two sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.
Sterilisation device/container dimensions: (L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15 mm), (L60 mm x B40 mm x H20 mm), (L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34.5 cm.
External measurement are preferably the same for both of the surgical departments but is a challenge due to the container size.
Operating theatre 1 D2.4052:
Maximum space is:
Depth: enter the outer measurement depth (must be able to take container L70 mm x B28 mm x H20 mm, L60 mm x B28 mm x H13 mm), (L54B x B26 mm x 15 mm).
Width: 900 mm-1 000 mm
Height: Can be a little higher than 1 970 mm
Operating theatre 2 D3.4023:
Maximum space is:
External measurements:
Depth: 1 020 mm (must be able to sterilise the sterilisation device (must be able to hold the sterilisation device: (L58 mm x B28 mm x H21 mm), (L62 mm x B42 mm x H20 mm)/L60 mm x B20 mm x H20 mm)
Width: 900 mm-950 mm
Height: can be a little higher than 1 970
The steam steriliser delivered must be able to be transported into the named sterilisation room without building intervention (structural works). OUS will not perform any alteration of doors/transport routes in the corridors to get the sterilisers into the sterilisation room, due to patient activity.
Transport route:
Door into the sterilisation room
Opr. 2 : 87.5 cm x 203.5 cm
Operating theatre 1 : 98 cm x 209 cm
One extra steam steriliser at Rikshospitalet [the National Hospital] needs to be replaced and we would like a price for one device as an option.