OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Diagnostic and interventional angiography KITOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
GUIDE CATHETER: high support with proximal reinforcement with flat section steel wires and round section nitinolo wires. Reinforcement of the distal part by means of nitinol spiraling.OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
5F, 6F and 7F guide catheters for extracranial coaxial catheterization and epiaortic trunks and for the release of interventional devices characterized by: 5F, 6F and 7F guide catheter for cathetOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Distal access guide catheter of external diameter 6F characterized by:Distal access guide catheter of external diameter 6F characterized by:OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Support guide catheter with nitinolo reinforcement with variable rigidity. Flexible distal end length of about 8cm. Distal marker. Available in two calibers:Support guide catheter withOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Guide catheter with distal balloon about 12-14 mm long, between two markers, for blocking blood flow, armed both internally and externally. External diameter 6 and 8Fr, length from 10 cmOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Long Armed Introducer 6Fr with large internal lumen preferably 0.088", available in 2 lengths 80, 90 cm approximately, and two different straight and multipurpose configurations. Flexible distal zoneOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Long vascular introducer with detachable valveOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Hydrophilic polymer guide - Diameter: 0.035"/0.038" - Length: 180cm, 260cm - Softness: standard, stiff - Tip: angledHydrophilic polymer guide - Diameter: 0.035"/0.038" - Length: 180cm,OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Neurovascular microguides for intracranial navigation, in steel, with a single element stem. Available with hydrophilic and polymer coating. Double coil in platinum and steel, with passageOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Stainless steel microguide with tapered gauge: preferably 0.008" distal and 0.012" proximal. Equipped with a preformable tip, it has a platinum coating in the distal tract of alOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Y-connectors with silicone valve and "push" opening and closing mechanism, Internal lumen: 7F, 20cm long side connectionOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Stainless steel microguides with triple hydrophilic coating with an extremely soft tip, coated in platinum in the distal section (at least 5 cm). Single element core in steel,OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Hybrid superselective microguide with steel and nitinol core, 14 mm hydrophilic preformable straight tip (about 40 cm distal) and radiopaque with platinum spiral, preferably 0.007" gauges, preferably 0.007",OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Super selective hybrid microguide with steel and nitinol core, preformable straight tip (about 14 mm) hydrophilic and radiopaque with platinum spiral, preferably caliber of 0.014" proximal and 0OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Microguide in superalloy alloy, in nickel-cobalt-chrome, preferably 0.014 "diameter of 0.014", distal spiral in radiopaque platinum, hydrophilic coating of about 36 cm. Perforated proximal end pOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Microguide with steel core gradually tapered with platinum wire that wraps the distal part of the inner core for greater radiopacity. The tapered part must have a nitin hypotubeOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Reinforced Nitinol microcatheter, equipped with high smoothness hydrophilic coating with double distal marker at 3 cm. Microcatheter body in Nyilon and Pebax type copolymer, internal lumen in PTFE.OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Precurved microcatheter with braided proximal portion and distal portion with reinforced spiral. In nylon, length at least 150 cm. Shapes: straight, 45°, 90° and J. Outer/distal diameter 2,3/1,7OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Hybrid microcatheter, with proximal armor consisting of a stainless steel hypotube of about 82cm while the distal part is an interweaving of filaments preferably of nitinol. 150cm longOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Reinforced stainless steel microcatheter, equipped with high smoothness hydrophilic coating. It has distal markers. Microcatheter body in Pebax type copolymer, internal lumen in PTFE. PointOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Flow-dependent microcatheter with hydrophilic coating, compatible with embolizing liquids. Nylon microcatheter body with distal reinforcement reinforcement in Nitinol, while the next partOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Percutaneous closure system of fenoral arterial access, with sandwich system formed by intravasal anchor and extravasal collagen sponge, both completely absorbable - DiametersOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Single lumen microcatheter with variable stiffness, hydrophilic surface and PTFE inner lining. Distal tip conformable to steam (by means of a shaping core, supplied together with theOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Microcatheters with double armature compatible with spirals of various sizes and curves. Measures preferably 1.9-3.8", about 155 cm.OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Microcatheters for flow navigation with hydrophilic coating external diameter preferably of 1.2-1.5-1.8 Fr distally flexible, compatible with guides preferably up to 0.08" and withOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Dedicated microcatheter for the release of 0.020" spirals, with an internal lumen of not less than 0.025". Proximal outer diameter preferably 2.95 F and distal outer diameter preferably dOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Tungsten-type radiopaque material to be mixed with liquid embolizersOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Platinum alloy vascular occlusion device with instantaneous mechanical controlled detachment. Consisting of a spiral of primary diameter preferably 0.020", shape and degree memoryOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Self-expandable device of ovoid shape in nitiniol, designed for vascular occlusion. It must have at least two platinum markers at both ends. The device must incorporate a revOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Compliant/extracompliant balloon catheter, double lumen: one independent for microguide preferably 0.0165'' and the other coaxial for inflation/deflation preferably 0.030OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Kit composed of balloon microcatheter with coaxial lumen complicating for temporary occlusion of cerebral vessels, and guidance preferably from 0.010 "Body of the microcatheter in Peba type copolymerOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Balloon mounted on monorail rapid exchange catheter for cerebral artery angioplasty. Catheter with hydrophilic distal segment of 40 cm with double lumen, total length 160 cm, diameters fromOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Y-shaped hemostatic valve, with five-leaf menbrane for optimal seal during angioplasty procedures. The device must be ergonomic and easily usable with one hand thanks also toOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Kit composed of supercompliant balloon microcatheter with coaxial lumen for temporary occlusion of cerebral vessels, and guidance preferably from 0.010 Body of the microcatheter in copolymer typeOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Low profile monorail (rapid exchange) type balloon catheters. Various diameters and lengths. Compatible catheters with 0.014” guide and a length of at least 130 cmOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Self-expandable stent for carotid use, built in nitinol, preassembled inside a monorail catheter, with conical and cylindrical configuration, of various sizes of diameter and length. CompatiOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Self-expanding stent for carotid use, made of steel, pre-mounted inside a monorail catheter, of various sizes of diameter and length. RM compatibleOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Self-expanding stent for carotid use, made of nitinol with double layer or with additional coatings, which guarantee a high coverage index of the plaque and the vessel wall to theOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Catheter with a large suction lumen (preferably 0.060", 0.064", 0.068"), length about 132 cm, proximal steel reinforcement and distal nitinol with progressive flexibility. Compatible with pomOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Thrombosuction catheter, double-layer reinforced structure: steel spiral and metal mesh, 6 fr, preferably 0.070 inch, lengths from 125 to 131 cm approx., hydro coatingOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Thrombi suction catheter with distal end flexible, hydrophilic, reinforced with helical spiral, outer diameter 6fr inside preferably 0.072"OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Intravascular foreign body recovery system consisting of loop or bascket or forceps, capable of navigating distally in intracranial vessels, length not less than 190 cm.OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Release controlled release embolization spiral in platinum alloy (92% platinum-8% tungsten)*** with instantaneous actuation manual detachment system (supplied separately for free use). The sOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
1000 cc transparent gland with analog pressure control that guarantees pressures from 300 mmHg upwards, made of durable fabric or plastic materialOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
High filling spiral with controlled detachment in platinum alloy with primary diameter 0.020" usable for large aneurysms and releasable with dedicated catheter. Equipped with sistOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Platinum-tungsten spiral*** with thermal detachment, available in various geometric versions type complex and helical, standard, soft, supersoft configuration can be used for filling and finishingOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Platinum spirals with electrolytic detachment through a dedicated device to be supplied free of charge, with isodiametric pusher, preferably measures from 1 to 24 mm in diameter, from 1 to 50 cm dOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Embolization spirals in platinum alloy, with instantaneous controlled detachment system, with detacher to be supplied free of charge. Stretch resistent and progressive softness. Length up toOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Release embolization spiral in platinum alloy (92% platinum-8% tungsten) ***with manual detachment system to be supplied in free use, with instant actuation. Spiral of emboliOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Spirals with variable conformation (both classic round and triangular) in platinum with monofilament in Prolene 10-0 inserted inside the spiral or with polyglycolic acid (PGA) inserted in theOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Self-expandable intracranial stent for small nitinol vessels with a diameter from 2.25 to 3.25mm and lengths preferably from 10 mm up to 25mm. Self-expandable intracranial stent for small vesselsOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Intrasaccular embolization device for intracranial aneurysms, consisting of a high-density Nitinol thread intertwine, self-expanding, compatible with microcatheters of internal caliberOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Stent for nitinol bifurcation aneurysms with 6 distal petals to create floor on the collar, flexible crown, 6 distal markers, electrolyte detachment. Crown sizes from 5 to 15 mm ciOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Stent for bifurcation aneurysms with flow diversion, in nitinol, with 4 distal petals covered by a non-permeable membrane perforable with micro for spirals. Compatible preferably with micOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Drains longer than 150 cm with high pressure filtered wellOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Self-expandable intracranial stent in braided Nitinol, diameter from 2 to 5.5 mm, lengths preferably from 12 to 75 mm. Platinum radiopaque propellers for better visibility. AvailableOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Nitinol stent for self-expandable and closed-cell neurovascular remodelling, fully retractable. The stent is attached to a steel pusher and the detachment is electrolyticDesign preferOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Nitinol stents for the treatment of wide-collared intracranial aneurysms. Self-expandable open-cell laser cut stent with countersunk proximal and distal part for better adaptationOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
System for remodelling of intracranial aneurysms that does not result in blockage of blood flow. Consisting of a nitinol mesh, totally radiopaque, and equipped with an opening and closing systemOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
System for remodelling of intracranial aneurysms that does not determine the blockage of blood flow.consisting of 42 total wires equipped with opening and closing system various sizes and compatible withOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Device for the removal of thrombi composed of an internal diameter stent of about 1.25mm and an external open-mesh stent of about 5mm diameter with at least two lengths, usable in vaeOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Fully retractable closed-cell nitinol device, stent treiver type, for the mechanical removal of thrombi from an intracranial blood vessel. The stent is attached to an ac pusherOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Intraluminal thrombus removal system composed of a fully retractable self-expandable nitinol structure with three-dimensional configuration. Available in length from 26 mm andOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
System for mechanical thrombectomy for the treatment of ischemic stroke. Nitinol mesh, totally radiopaque, with controllable expansion through a fixed handle that allows the opening andOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Stent for recovery of intracranial thrombi in self-spreading nitinol, closed cells, closed proximal ring. Version also completely radiopaque, or with proximal and 2 distal markers. Stent for recOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Neuroradiological diagnostic catheter - Outer Diameter: 4F, 5F, 5.5F, 6.5F - Length: 100cm and 125cm - Curves: Pig Tail, HH1, JB1, Simmons1, Simmons2, Vert, Neuroradiological diagnostic catheter - DOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Non-detachable stent-shaped intracranial thrombi recovery system, in self-expending Nitinol completely radiopaque and totally visible in scopy, equipped with at least 2 or 3 radiopaque markersOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Intracranial small-vessel flow diversion stent navigating within a 0.17 microcatheter Intracranial flow diversion stent for small vessels navigating within aOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Stent flow diverter in nitinol consisting of 64 and 48 wires, completely recaptureable with voluntary detachment and version with independent guide pusher with the possibility of advancement, withdrawal and rotasOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Self-expandable flow diversion device with very dense mesh, composed of 48 intertwined microfilaments of which 3/4 in Chrome / Cobalt and the remaining 1/4 in Platinum / Tungsten, packaged in aOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Self-expanding stent consisting of 48 intertwined niquinol microfilaments with platinum inner core for greater radiopacity, equipped with proximal and distal markers. Up to 90% ofOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
KIT for spinal interventional intervention consisting of:OPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Bone biopsy kit consisting of a graduated T-needle of about 10 gauges, made of steel with spindle and coaxial device to take the sample inside the boneBone biopsy kit consisting ofOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Intracranial access system made of NYLON, PEBAX type copolymers, and URETHANE (with stainless steel reinforcement), internal coating in Ptfe available in 2 different versions, 6 F proximalOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY
Intracranial distal access system, tapered with proximal support portion 6F of monolumen catheter of variable rigidity with nitinol reinforcement spiral. Equipped with a radiopaque markerOPEN PROCEDURE FOR THE AWARD OF THE SUPPLY OF MEDICAL DEVICES FOR INTERVENTIONAL NEURORADIOLOGY