Miscellaneous medical devices and products

Contract Value:
EUR 4M - 4M
Notice Type:
Contract Notice
Published Date:
06 April 2022
Closing Date:
23 May 2022
Location(s):
ITI43 Roma (IT Italy/ITALIA)
Description:
supply of medical devices for electrostimulation/pacing, extraction of infected lead catheters and electrophysiology divided into 46 lots,

Open procedure telematic tender within the Community, pursuant to art. 60 of Legislative Decree 50/2016 and subsequent amendments, for the conclusion of a framework agreement with multiple economic operators pursuant to art. 54, paragraph 4, lett. a) of Legislative Decree 50/2016 for the supply of medical devices for electrostimulation/pacing, extraction of infected lead catheters and electrophysiology divided into 46 lots, needed by the hospital company- Sant'Andrea University in Rome for a period of 3 (three) years.

CIG 9097211800

Entirely subcutaneous implantable defibrillators with its defibrillation lead. Essential characteristics that the device must possess: at least 3 programmable sensing vectors; MRI compatibility; advanced AF diagnostics.

CIG 9097993D52

Absorbable antibacterial shell for the prevention of infections by implantation of devices for cardiac electrostimulation. The device must have the following essential characteristics: Local polymer-controlled release of at least two antibiotic molecules.

CIG 909800680E

Long manual extraction cannulas. The device must have the following essential characteristics: Cannula entirely visible during fluoroscopy

CIG 9098015F79

Stiletto for lead traction and extension and anchoring accessories.

CIG 909808486C

Rotating dissector for the mechanical removal of cardiac leads for stimulation and defibrillation. The device must have the following essential characteristics: Device consisting of a handpiece and a dissection blade inserted in the distal part of the device

CIG 9098093FD7

Balloon catheter for temporary occlusion of the superior vena cava, complete with preparatory kit.

CIG 9098114130

Sheaths for excimer laser removal of cardiac leads for stimulation and defibrillation. The winning company will have to make the excimer laser generator available on free loan for use, with the burden of maintenance and any consumables. Laser sheath in a single sterile package. The device must have the following essential characteristics: Different sizes of laser sheaths

CIG 9098126B14

Kit for removing electrodes and catheters with or without a free head by transfemoral route. The device must have the following essential characteristics: Ability to grasp intracavitary objects with or without a free head

CIG 90981330DE

Quadripolar, hexapular, octopular, decapular diagnostic lead leads for electrophysiological studies, with FIXED curve, in polyurethane, with related connection cables. The device must have the following essential characteristics: Minimum diameter 6 French; Platinum and/or platinum-iridium electrodes

CIG 90981341B1

QUADRIPOLAR, HEXAPOLAR, OCTOPOLAR, DECAPULAR ADJUSTABLE diagnostic lead leads, for electrophysiological studies in polyurethane with related connection cables The device must have the following essential features: minimum diameter 7 French, Push-Pull or Rotational piston handpiece.

CIG 9098141776

ADJUSTABLE diagnostic duodecapular lead with relative connection cables. The device must have the following essential characteristics: Minimum diameter 7F, Body in thermoplastic elastomer or polyurethane, Mechanically oriented clutch mechanism, Availability of different types of curves.

CIG 90972464E3

SINGLE-CHAMBER leadless pacemaker complete with release kit and introducers. The device must have the following essential characteristics: SSI stimulation mode, SSIR; programmable base stimulation frequency from 40 to 120 bpm; Output ≥ 5 V amplitude output and ≥ 1 ms of lifespan

CIG 9098150EE1

ADJUSTABLE multipolar lead for mapping the tricuspid ring and right atrium, availability of at least 20 poles and various spacings.The device must have the following characteristics essential: Minimum diameter 7F, body in thermoplastic elastomer or pebax, mechanism with bi-directional or push and pull mechanically oriented clutch.

CIG 9100596166

Circular, multipolar lead for the mapping of pulmonary veins. The device must have the following essential characteristics: Minimum diameter 7F, Variable diameter model, Mechanism bidirectional micrometric mechanical orientation or push and pull

CIG 9100601585

Multipolar lead for high-density mapping, active curve, equipped with at least 16 electrodes, compatible and viewable with an electroanatomical reconstruction system, with related connection cables. The device must have the following essential characteristics: Mechanical, micrometric, push and pull mechanism, or rocker

CIG 91006058D1

Electrocatheters for endocardial ablation, adjustable radiofrequency tip 4mm, 8 mm, pebax body with relative connection cables. It requires radio frequency generator and plates necessary for the use of the hectrocatheter. The device must have the following essential characteristics: Ergonomic handpiece with adjustment and locking mechanism, possibility of different types of curves, thermistor or thermocouple system.

CIG 9100612E96

Radiofrequency endocardial ablation leads with 8 mm tip with advanced capabilities to recognize intracavitary signals with related connection cables. It is required on free loan for use, radio frequency generator and plates necessary for the use of the hectrocatheter. The device must have the following essential characteristics: Ergonomic handpiece with micrometric bidirectional adjustment mechanism with automatic lock, Possibility of different configurations of the distal part with various lengths and different types of curves, Availability of 3 120° mini-electrodes integrated on the tip for pacing and EGM reading, Pebax body

CIG 91006161E7

Radiofrequency irrigated scaler lead, active curve, with the presence of a magnetic sensor at the tip, compatible with electroanatomical reconstruction systems. The company must provide, on free loan for use, the radio frequency generator, plates necessary for the use of the hectrocatheter and peristaltic pump. The device must have the following essential characteristics: Push and pull type micrometric handpiece, or rocker type, Possibility of different curves, Body of the lead in polyurethane or pebax

CIG 910062487F

Scaler lead with contact or force sensor, radio frequency with cooled tip of 3.5 mm or 4 mm or 4.5 mm, multiple holes, compatible with electromagnetic reconstruction system and related cables of connection. The company must provide the radio frequency generator, peristaltic pump and plates necessary for the use of the lead free of charge. The device must have the following essential characteristics: Micrometric push and pull handpiece, or rocker type, Possibility of different curves.

CIG 9100629C9E

Focal cryoablation leads, quadripolar, with 4 and 8 mm tip. The company will have to provide a cryoablation system on loan for use. The device must have the following essential characteristics: Availability of different bending radii

CIG 9100636268

Balloon adjustable catheters for cryoablation of pulmonary veins, including connection cables, deflectible introducer and circular catheter for mapping pulmonary veins. The company will have to provide a cryoablation system on loan for use

CIG 91006405B4

Intracardiac ultrasound probe that can be integrated with a non-fluoroscopic mapping system. The company will have to provide an ultrasound machine on loan for use. The device must have the following essential characteristics: Maximum diameter 10 F, Display of endocardial structures in high resolution, Working frequency between 3 and 10 MHz, Possibility of connection of the probe to an electro-anatomical mapping system.

CIG 9097267637

Leadless pacemaker with the possibility of atrial sensing complete with release kits and introducers. The device must have the following essential characteristics: SSI stimulation mode, SSIR, DDI, DDIR; programmable stimulation base frequency from 40 to 120 bpm; Output ≥ 5 V amplitude output and ≥ 1 ms duration

CIG 9100650DF2

Transeptal puncture kit: introducer and needle. The device must have the following essential characteristics: minimum diameter 8F, minimum needle length 71 cm

CIG 9100659562

Steerable introducers for the positioning of lead catheters for transeptal access to the vascular system. The device must have the following essential characteristics: Maximum diameter 11F, Availability of different types of curves, Complete with dilator, valve.

CIG 9100 669 FROM 0

Transeptal puncture needles compatible with a steerable introducer for transeptal ablation. The device must have the following essential characteristics: Minimum length 71 cm, stainless steel needle

CIG 910067743D

J-Tip Transeptal Puncture Guidewire The device must have the following essential characteristics: Diameter 0.014", Length 135cm

CIG 9100686BA8

High density mapping lead with contact technology and magnetic sensor compatible with the electromagnetic mapping system of heart chambers. It is required to supply the mapping system for the use of the diagnostic catheter and related connection cables. The device must have the following essential characteristics: Mapping lead with minimum number of 20 poles, Max 8.5 F catheter body, Magnetic sensor for system compatibility of mapping.

CIG 9100694245

Bipolar leads for temporary stimulation

CIG 9100700737

External reference patch system to be applied to the patient for compatibility with the electromagnetic mapping and ablation system. It is required to supply the mapping system for the use of patches and related connection systems. The device must have the following essential characteristics: Compatibility of the patch system and electromagnetic field with high density diagnostic catheter

CIG 9100705B56

External reference patch system to be applied to the patient, integrated with an electromagnetic mapping system. The supply of the mapping system for the use of the Patch and related connection systems is required

CIG 910071646C

External reference patch system to be applied to the patient, integrated with an electromagnetic mapping system, preferably without the need to use dedicated catheters. The supply of the mapping system for the use of Patches and related connection systems is required

CIG 9100734347

Esophageal lead for esophageal temperature monitoring. Temperature control device and supplied connection cables are required. The device must have the following essential features: Availability of at least 7 sensors for temperature monitoring on the catheter

CIG 909728985E

Injectable, implantable cardiac monitoring devices, complete with remote monitoring. The device must have the following essential characteristics: injectable device, MRI compatibility, presence of programmable recording modes dedicated to the management of arrhythmias; inclusive of remote monitoring system

CIG 910074190C

Non-invasive arrhythmia mapping system consisting of a jacket complete with a work station for data processing. The successful bidder must provide, on free loan, a work station for the processing of data acquired by the device. The device must have the following essential characteristics: Patient wearable vest equipped with electrodes for the reconstruction of electroanatomical maps, TAC image integration system

CIG 910075114F

Peel-away introducers. The device must have the following essential characteristics: Complete kit with 10 ml syringe, needle for central vascular access, guide >0.34 inch for introduction according to the Seldinger technique

CIG 91007597E7

External defibrillation plates. The device must have the following essential characteristics: Supplied with external, radiolucent connection cables

CIG 9100818897

Biotome for endomyocardial biopsies. The device must have the following essential characteristics: Presence of dedicated introducers, for femoral access length at least 100 cm - for jugular access length at least 45 cm.

CIG 910083569F

Complete pericadiocentesis kit for pericardial suction/drainage. The device must have the following essential characteristics: Complete Pericardial Aspiration/Drainage Kit

CIG 9100861C12

Sterile procedure kit for operating table for interventional cardiology procedures. The device must have the following essential characteristics: 200x230 mother towel with super-absorbent inner layer and waterproof outer layer, Two sterile surgical gowns, packed XL with waterproof protection in TNT and superabsorbent outer layer and waterproof interior welded to the gown with double seams and edged flaps; back of the shirt with wallet closure, Cover for PMK implant and interventional cardiology procedures (260x350) in TNT with high absorbency outer layer with superabsorbent TNT reinforcement and presence of 4 holes with transparent protective film correspondence of the inguinal region and the infraclavicular parasternal region bilaterally, 4 bowls available in various diameters and colors, 5 syringes plus needles, scalpel, 2 sterile headsets for instrument covers available in various sizes 40 sterile gauze available in various sizes, 4 sponges for patient's skin disinfection.

CIG 9100866036

High-resolution intracardiac ultrasound probe. The company will have to provide an ultrasound machine on loan for use. The device must have the following essential characteristics: Maximum diameter 10 F, Display of endocardial structures in high resolution, Working frequency between 3 and 10 MHz,

CIG 9097893ACD

Wearable defibrillator (express rental requirement for no months per year)

CIG 9097913B4E

Implantable cardiac device for the treatment of patients with chronic heart failure, based on the delivery of cardiac contractility modulation therapy. The device must have the following essential characteristics: Biphasic pulse delivery (CCM therapy) during the absolute cardiac refractory period, cardiac device with rechargeable battery from external (dedicated charger); possibility of implantation in patients with atrial tachyarrhythmias

CIG 90979287B0

Implantable cardiac device for the treatment of patients with chronic heart failure, based on the delivery of cardiac contractility modulation therapy WITH THERAPY AVAILABILITY ANTI-ARRHYTHMIC, for the prevention of sudden death. The device must have the following essential characteristics: Biphasic pulse delivery (CCM therapy) during the absolute cardiac refractory period, cardiac device with rechargeable battery from external (dedicated charger); Device capable of delivering anti-tachycardic therapies

CIG 9097949904

Permanent electrostimulation system in kit for His Bundle Pacing. The device must possess the following essential characteristics: System of leads and introducers for selective stimulation of the His beam

CIG 9097959147

Sterile cables for measuring electrical parameters during the implantation of pacemakers or defibrillators. The device must have the following essential characteristics: Pin connection or alternatively equipped with a PSA adapter

Download full details as .pdf
The Buyer:
Azienda Ospedaliero Universitaria Sant’Andrea
CPV Code(s):
33190000 - Miscellaneous medical devices and products