A Water treatment system is required for the satellite dialysis unit at Bognor Dialysis Unit and is required to supply ultrapure reverse osmosis quality water to the renal ward area. The system must be installed without disruption to the existing dialysis provision. The expectation therefore, is that this must be a separate suite of plant together with the associated RO loop etc is the scope of supply.The tender response for the water treatment system must include both pre-treatment and reverse osmosis as a combined proposal.The installation of the water treatment system must conform to current National and International guidance as listed below:Current Guidance:1. BS EN ISO 23500-1:2019 : Preparation and quality management of fluids for haemodialysis and related therapies. General requirements.2. BS EN ISO 23500-2:2019 : Preparation and quality management of fluids for haemodialysis and related therapies. Water treatment equipment for haemodialysis applications and related
For the purchase of Water Plants at Bognor and IOW Dialysis Units