Inspections and repairs of medical equipment
System reviews for non-invasive diagnostics of the circulatory and autonomic systemsAccording to SWZ records and price form
Spare parts for review:
- Pompa ContBP 12V
- Pompa OscBP 6V
- Elektrozawór OscBP
Inspections:
- Regulacja ECG, ICG, OscBP, ContBP
- Kalibracja ECG, ICG, OscBP, ContBP
-Disinfection
- Electrical safety tests
- Leak tests
Inspections of GDX-70/C2H4O measuring heads (with replacement equipment)According to SWZ records and price form
- Dismantling the head
- Installation of a replacement head
- Calibration in the measurement laboratory
- Dismantling the replacement head
- Head assembly
Replacement equipment included in the price of the inspection
Inspections of the cryo-surgery apparatus CRYO-S ELECTRIC II and the cryosurgery apparatus CRYO-S CALSSICAccording to SWZ records and price form
- General assessment of the condition of the camera
- Cleaning the device
- Functionality tests
- Replacement of O-rings and leak check
- Verification of measuring instruments of the device
- Electrical safety tests
X-ray diagnostic monitor reviewsAccording to SWZ records and price form
- External control of the device
- Safety measurement with the right meter
- Replacement of bactericidal radiator
- Cleaning and maintenance
- Measurement of functional parameters
- Monitor calibration
Inspections of the VNG device, the ENTHERMO 304T water caloric simulator and the SMART EP ARRS OAE auditory evoked potentials test deviceAccording to SWZ records and price form
Annual review including:
- Checking the technical condition
- Validation of work
- kalibrację
Electrocardiograph reviews ascard green v.06.101According to SWZ records and price form
Annual review, which includes:
- Electrical safety test in accordance with IEC 60601-1 (meeting the safety standard PN EN 62353) carried out with a dedicated Rigel 288 tester.
- Control the condition of the case, the operation of the keyboard, printer and fastening of elements.
- Functional test carried out with the Rigel Uni-Pulse 400 tester (ECG simulator) and a test confirming the correct operation of all functions in accordance with the technical documentation.
Electrocardiograph reviews LUMED EUROECG with trolleyAccording to SWZ records and price form
Technical Inspection
-Maintenance
- Electrical safety tests
- Checking the operation of the camera for WTO compliance 25/2020
- Necessary instruments:
* Electrical safety tester RIGEL 277
* RIGIEL 33 Simulator
* BRYMEN digital meter
Surgical laser reviews VELAS PRO-30B15DAccording to SWZ records and price form
- Electrical safety check
- Checking the level of laser radiation emission
Inspections of the NWM monitoring apparatus and diagnosis of breathing disorders during sleepAccording to SWZ records and price form
Annual review, which includes:
- Visual inspection, checking the serial numbers of components.
- Verification of sensors / all system components / verification of correct operation of channels (impedance)
- Checking the video path
- Software update/replacement
- Consumable check, condition and expiration date
- Performing a test test.
A-SCAN PACSCAN 300 Ophthalmic Ultrasound Reviews - BIOMETERAccording to SWZ records and price form
- Electrical safety tests
- Function test
- Visual assessment
- Checking the correctness of operation
FISH System OverviewAccording to SWZ records and price form
Annual review, which includes:
- cleaning of optical surfaces (lenses, glasses, spectacle cap, condenser, glass plate - at field aperture, available filters, polarizing elements)
- cleaning of mechanical surfaces (table tripod, handle of preparations, caps),
- Koehler lighting setting,
- setting polarizing filters,
- checking the correctness of the electrical system,
- regulation of the video path,
- checking the centering mechanism in the universal condenser,
control and possible lubrication of table guides.
Reviews of GENERAL MIDITECH transport cardiac monitorsAccording to SWZ records and price form
- Technical review
- Technical maintenance \measurements and functional tests
- Tests and calibration
- Electrical safety test
- Control of operating parameters
Inspections of the VIVID IQ ultrasound echocardiography machineAccording to SWZ records and price form
- Camera maintenance
- Calibration of transducers
-Electrical safety tests according to IEC 606001-1
Przeglądy diatermii AUTOCON III 400According to SWZ records and price form
-Cleaning
- Power measurement
- Functional tests
- Electrical safety tests
Inspections of the MESOTEST II nyctometerAccording to SWZ records and price form
- Odkurzenie
- Washing optics
- Checking the correctness of operation
- Check connections
- Software update (if newer version is available)
- Camera calibration
- Electrical safety test according to 62353
Reviews of the device for controlled hypothermia CRITICOOLAccording to SWZ records and price form
- Visual inspection
- Functional tests
- Control of operating parameters
- Filter replacement
Inspections of washer-disinfector type PWD 82545 SADAccording to SWZ records and price form
- Leak check
* Water inflow/outflow
* Dosing system
* Pressure booster pump
* Door, door gasket
- Visual inspection and operation
* Washing arm: pan/bolt function control
* Basket insert: control of correct locking
* Inspection of washing nozzles
* Tank: float switch control, cleaning if necessary
* Water drainage check
* Door lock control
* Rollover protection
- Dosing system control
* Check dispensing systems, including emptying message
* Optical dosing control in tank and/or boiler
* Dosing calibration
- Dosing/calibration of process chemicals
* Stone binder
*Detergents
- Security checks
* Thermoelectric measurement
* Security control
- Parameter changes
Przeglądy kardiomonitorów Cardio 9000 STANDARDAccording to SWZ records and price form
- full control of the safety of the cardiac monitor,
- full control of cardiac monitor operating parameters,
- visual inspection of the cardiac monitor:
* assessment of the visual condition,
* assessment of the condition of the wires,
* assessment of the earthing condition,
* assessment of the condition of equipment,
- cleaning the cardiac monitor and internal components of the device from dust and dirt,
- service calibrations if calibration is necessary:
* Capnography sensor
* NIBP module
- functional tests,
- service tests (if the software provides for the possibility of performing tests or the software is not blocked),
- control of operating parameters using control and measurement equipment (ECG, SpO2, NIBP, TEMP, IBP, CO2, CO, ICG),
- electrical safety test,
- Reset the alert regarding the required inspection and programmatically set the date of the next review
Inspections of the NEOPROBE 2300 NPCU23 gamma radiation detector and SCM23 mammotomeAccording to SWZ records and price form
Replacement equipment for the time of inspection included in the price of the inspection
Detector:
- Checking accessories
- Cleaning the unit
- Functional test
- Electrical safety test
- Test according to instruction number 00-0298
Mammotom:
- Cleaning the unit
- Functional test
- Electrical safety test
- Checking accessories
Inspections of the THOPAZ digital chest drainage systemAccording to SWZ records and price form
- Control and visual verification of the device
* Verification of the external condition of the device - checking the occurrence of: cracks, mechanical damage, discoloration on the housing and manometer
* Inspection of the presence of technical socket covers (power supply and data ramping)
* Checking the correct operation of the mounting brackets
* Readability of the nameplate
* Control of the efficiency of the canister release button
* Checking the presence of canister gasket
* Device cleanliness rating
* Efficiency of the docking station and power supply
- Device tests
* Making an FPT test with a measuring pressure gauge
* Conducting a functional test
* Checking the correct operation of battery power
* Checking the acoustic signal (alarms)
* Battery charging test
- Spare parts available
* Dren
*Can
* Uszczelka
- Electrical safety test standard 62353
Przeglądy kabiny z promiennikami emitującymi UVB/UVA WALDMANN UV 7002According to SWZ records and price form
Annual review, which includes:
- Checking and calibrating the intensity of UVA and UVB radiation emitted by the cabin,
- Performing a functional test
- Taking possible measures to eliminate the problem (without the use of spare parts)
- Electrical safety test
- Every 24 months replacement of air filters in the UV7002 cabin - 1 time during the term of the contract
Przeglądy kardiotokogramu BTL FCAccording to SWZ records and price form
- Checking the technical condition
- Electrical safety tests
Inspections: ABATOR 50AE, AERATOR XL, EO gas sterilizer with equipment and 5XL gas sterilizerAccording to SWZ records and price form
Sterylizator:
- Review every 6 months
- Dismantling of compressed air filters, replacement of 0.3 micron and 0.1 micron filter with O-rings
- Checking the ventilation line
* Check for tightness of connections between the sterilizer and the installation in the hospital
* Check for bends and other blockages that may interfere with proper air circulation
* Checking if the installation is done according to the specification
Humidification system
- Replacement of the stem and O-rings of the humidification valve; if necessary, replacement of the solenoid valve
Vacuum system
- Cleaning the air solenoid valve stem. Replacement if necessary
- Replacement of air reducer components using a repair kit
- Vacuum pump cleaning
- Door locking system
* Cleaning the door closing solenoid valve. Cleaning the stem and socket; replacement if necessary
* Control of the door lock mechanism
- Vent system
* Disassembly of the breather solenoid valve. Cleaning the stem and socket; replacement if necessary.
* Replacement of the bacterial filter
- Dismantling of the punch solenoid valve. Replacement of mandrel and O-rings.
- Cleaning of the water system
* Dismantling of the water tank
* Rinse with clean water
* Checking the PVC pipe connecting the tank with the solenoid valve for cracking; if necessary, replacement with a silicone tube of the same diameter
Abattoir:
- Abator filter replacement
- Air flow measurement (fan check)
- Catalyst temperature check
Aerator:
- Replacement of the bacterial filter - 2 pcs.
Autoefractometer inspections with non-contact tonometer function TONOREF II NIDEKAccording to SWZ records and price form
AUTOREFRACTOMETER WITH NON-CONTACT TONOREF II NIDEK FUNCTION
- external visual inspection;
- control of the nameplate;
- housing wear assessment;
- tightening the screws fixing the housing;
- control of the operation of buttons;
- power outlet control;
- electrical safety test according to PN:EN 62 353
- control on work using an artificial eye
Inspections of ultrasonic knife PIEZOMED SA-320According to SWZ records and price form
A review once every 3 years, which includes:
- Check for errors in the device's memory.
- Verification of all functions.
- Verification of the correctness of the piezoelectric transducer.
- Software upgrade (if available).
- Check the condition of the water pump rollers.
- Maintenance of connections on the electronics board.
Harmonic knife inspections GEN11 (with replacement equipment)According to SWZ records and price form
-Disinfection
- Visual inspection
- Functional test
- Software update
- Output verification
- Electrical safety test
- Quality assurance test
Replacement equipment for the time of inspection included in the price of the inspection
Przeglądy kriokomory ALASKA AL01According to SWZ records and price form
- Checking the condition of the outer and inner cryochamber plating
- Checking the condition of glass and seals
- Checking the condition of the door: the treatment chamber and the entrance to the vestibule 1 and 2
- Checking the condition of floor panels
- Checking the tightness of the power supply system
- Checking the patency of the power supply system and the condition of the nozzles
- Checking the chamber supply system with cryogenic liquid
- Checking the cryochamber drying system
- Control system check
- Control software check
- Checking the condition of lighting: external cryochamber, treatment chamber, atria
- Checking the status of door lock sensors
- Checking the light condition of the visualization of the location of the doors of the vestibules 1 and 2 and the treatment chamber
- Checking temperature sensors
- Checking the functioning of the controller
- Checking the functioning of the T1 and T2 temperature measurement system in the treatment chamber
- Check of the emergency shutdown system (safety)
- Checking the condition of the panel with filters for dust removal from ice dust cryochamber
- Checking the condition of the blinds between the accumulation chamber and the treatment chamber
- Check the status of the emergency exit panel
- Checking the functioning of the voice communication system with the patient
Inspections of syringe infusion pumps SP-12 SPRO-NAccording to SWZ records and price form
- Checking the correctness of operation
- Technical maintenance
- Measurements and functional tests
- Tests and calibrations
- Electrical safety test
Inspections of the OCULIGHT GL IRIDEX Ophthalmic LaserAccording to SWZ records and price form
- Checking the laser beam power with a dedicated meter
- Electrical safety test in accordance with PN-EN 62353
Skull trepan repairsAccording to SWZ records and price form
1 - skull trepan GB304R 12/15 mm with HUDSON tip - GB 304 R - 051097
2 - TREPAN SKULL - GB 304 R - 049393
3 - TREPAN SKULL TE563 - GB 304 R - 062168
4 - skull trepan GB304R 12/15 mm with HUDSON tip - GB 304 R - 51098
5 - skull trepan GB304R 12/15 mm with HUDSON tip - GB 304 R - 51101
6 - trepan skull GB304R 12/15 mm with HUDSON tip - GB 304 R - 051077
7 - Skull bone perforator - GB 304 R - 047982
8 - skull trepan GB304R 12/15 mm with HUDSON GB 304 R 051076 tip
9 - trepan skull - GB 304 R - 049198
10 - skull trepan GB304R 12/15 mm with HUDSON tip - GB 304 R - 051085
11 - TREPAN SKULL - GB 304 R - 051520
12 - TREPAN SKULL - GB 304 R - 049703
13 - TREPAN SKULL - GB 304 R - 048201
14 - TREPAN SKULL - TE 563 - 063650
15 - TREPAN SKULL TE563 - TE 563 - 062168
16 - TREPAN SKULL TE563 - TE 563 - 063668
17 - TREPAN SKULL - TE 563 - 063679
18 - TREPAN SKULL - TE 563 - 063674
19 - trepan skull - TE563 - TE 563 - 062968
20 - trepan of the skull - TE 563 - 050361
21 - Trepan skull TE563 - TE 563 - 062242
22 - TREPAN SKULL - TE 563 - 050480
23 - TREPAN SKULL - TE 563 - 062669
24 - trepan skull GB304R, TE563 12/15MMTE - TE 563 - 050368
25 - trepan skull GB 300R - TE 563 - 036808
26 - skull trepan GB302R 9/12 mm with HUDSON tip - TE 561 - 063082
27 - skull trepan GB302R 9/12 mm with HUDSON tip - GB 302 R - 063124
28 - skull trepan GB302R 9/12 mm with HUDSON tip - GB 302 R - 063121
29 -Trepan skull GB302R - GB 302 R - 060671
30 - skull trepan GB302R 9/12 mm with HUDSON tip - GB 302 R - 063115
31 - TREPAN SKULL - GB 302 R - 062283
32 - skull trepan GB302R 9/12 mm with HUDSON tip - GB 302 R - 063118
33 - TREPAN SKULL - GB 302 R - 0644150
34 - TREPAN SKULL - GB 302 R - 062272
35 - trepan skull GB302R 9/12 mm with HUDSON tip - GB 302 R - 063084
36 - Climb 9x12 - GB 302 R - 064681
37 - skull trepan GB302R 9/12 mm with HUDSON tip - GB 302 R - 63099
38 - skull trepan GB302R 9/12 mm with HUDSON tip - GB 302 R - 63114
39 - TREPAN SKULL - GB 302 R - 053561
40 - Trepan skull TE562 - TE 562 - 062210
41 - TREPAN SKULL - TE 562 - 062807
42 - trepan skull GB302R - TE 562 - 063019
43 - trepan skull GB302R - TE 562 - 061615
44 - TREPAN SKULL - TE 562 - 061693
45 - TREPAN SKULL - TE 562 - 064513
CLEMENT-CLARK Synoptophore 2001 and CANON RK-F1 autorefractometerAccording to SWZ records and price form
- external visual inspection;
- control of the nameplate;
- housing wear assessment;
- tightening the screws fixing the housing;
- control of the operation of buttons;
- power outlet control;
- electrical safety test according to PN:EN 62 353
- in the case of autorefractometer, there will be an additional check on the work using an artificial eye;
Inspections of BTL 4625 SMART electrotherapy devicesAccording to SWZ records and price form
-Electrical safety tests
- Checking the correctness of operation
Reviews of BRAINLAB neuronavigation systemsAccording to SWZ records and price form
- Visual and mechanical inspection of all system hardware components and instruments (if available) according to Brainlab inspection protocols
- Comprehensive maintenance of the computer unit (replacement of filters, cleaning of the hard disk, installation of the latest security patches, etc.)
- Electrical safety test
- Verification of third-party integration (e.g. C-arms, microscopes, etc.)
In addition, with neuronavigation - 2021:
- Checking the accuracy of the optical tracking device (Vega camera)
Inspections of the apparatus for magnetic isolation of genetic material Maxwell RSC 48According to SWZ records and price form
Annual review of the Maxwell RSC 48 including:
- Full system control,
* Cleaning and inspection of all components,
* External enclosure cleaning and inspection
* Cleaning and inspection of the device platform and internal surfaces
- Control of adjustment and adjustment of the device
- Software update
Technical inspections of Bechtold ELTRON 80 surgical diathermy and Bechtold ELTRON 80 surgical diathermy with ELTRON SE smoke suction with trolleyAccording to SWZ records and price form
- Security and functionality tests
- Measuring equipment necessary to perform the inspection - CHIROTEST Measuring Tester
Inspections of RADICAL7 capnometers with ROOT with ISA/Nomoline MASIMO moduleAccording to SWZ records and price form
-Check
- Kalibracja
- Electrical safety test