Angiography supplies | Tenderlake

Angiography supplies

Contract Value:
-
Notice Type:
Contract Notice
Published Date:
13 September 2023
Closing Date:
10 October 2023
Location(s):
PL POLSKA (PL Poland/POLSKA)
Description:
Supply of equipment for endovascular treatments for the Hemodynamics Laboratory

Supply of equipment for endovascular treatments for the Hemodynamics Laboratory

Number of packages – 24

Package 1 – Semi-compliant balloons

No. Detailed assortment Unit of measurement Quantity per 24 months Net Price Gross Price Net Value Gross Value Manufacturer and catalog number Notes

1 Semi-compliant balloons 1 pc 500

TOGETHER:

- hydrophilic coating, resistant to scratches and damage when tightening the stent;

- system monorail;

- crossing profile ≤ 0.021 inches for balloon Ø 3.0 mm;

- entry profile ≤ 0.017 inches for balloon Ø 3.0 mm;

- nominal pressure 8 atm for all sizes;

- pressure RBP - 14 atm for all sizes;

- different balloon lengths from 6 - 30 mm; - various ;

- balloon diameters from 1.2 - 5.0 mm varying every 0.25 mm in the diameter range of 2.0 - 4.0 mm.

PACKAGE NO. 10 – Microcatheter for CTO

Lp Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Microcatheter for CTO - 20 pcs

TOGETHER:

Microcewer for CTO

- has a taped safflower with a proximal diameter of 2.8 F and a distal diameter of 2.6 F

- microcatheter reinforced with tungsten weave

- available in lengths of 135 cm and 150 cm

- inner diameter of the tip 0,015"

- inner diameter of the cabinet 0,018"

- compatible with 0.014" chain

- maximum pressure 300 psi

- has a polymer hydrophilic coating on the distal 60 cm cabinet

- has a soft, atraumatic and upholstered tip

Package No. 11 – Microcatheter for CTO

Lp Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1

Microcatheter for CTO - Pcs. 20

TOGETHER:

Microcewer for CTO

- microcatheter reinforced with braid from 14 wires

- tip is visible in Skopje thanks to tungsten (tungsten resin)

- has in the distal part between the saffter and the hub spiral protection against bending of the catheter

- available in lengths of 135 cm and 150 cm

- maximum pressure 300 psi

- compatible with 0.014" (0.36mm) chain

- outer diameter of the cabinet in the distal segment 0.71mm (2.1F)

- outer diameter of the cabinet in the proximal section 0.95mm (2.9F)

- inner diameter of the cabinet 0.48mm (0.019")

- tip – upholstered, soft, atraumatic

- inner diameter of tipu 0,38mm (0,015")

- input profile = outer diameter tipu 0.44mm (1.3F)

Package No. 12 – Microcathray for CTO

Lp Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Microcatheter for CTO - 20 pcs

TOGETHER:

Microcewer for CTO

- has a taped safflower with a proximal diameter of 2.6 F and a distal diameter of 1.9 F and a tip with a diameter of 1.4F

- internal channel covered with PTFE, diameters: distal 0.017", proximal 0.022"

- braid made of 18 steel wires

- available in lengths of 135 cm and 150 cm

- inner diameter of the tip 0,016"

- compatible with 0.014" chain

- has a polymer hydrophilic coating on the distal 70 cm cabinet (for a microcatheter with a length of 135 cm) and 85 cm (for a microcatheter with a length of 150 cm)

- has a soft, atraumatic and upholstered tip

- microcatheter tip clearly visible in Skopi due to the content of tungsten powder

PACKAGE NO. 13 – Software package for coronary reserve analysis based on coronary angiography

Lp Detailed assortment Unit of measurement Quantity Net price Gross price Net value Gross value Manufacturer and part number Notes

1. License 12 months without limitation of the number of analyzes. Installation and commissioning - 1pc

2. Workstation - 1pc

3. Medical monitor 24" - 1 pc

4. Non-medical monitor - 1pc

- description monitor, monitor size min.24 '' fullHD IPS LED matrix, video inputs: 1xDP, 1xDVI, 1xVGA, PIVOT function

TOGETHER:

Description of items to be valued, Quantity

1. License 12 months without limitation of the number of analyzes. Installation and commissioning

2. Workstation

3. 24" medical monitor

4. Monitor niemedyczny

- Viewing heart examinations in 2D mode, with the option of viewing in cine mode, presentation of reference lines,

- Ability to measure the distance and area of the selected segment,

- Dedicated to different types of heart pathology display presets of heart examinations,

- Possibility of importing AX, CT, MR, IVUS tests from CD/DVD and other data carriers,

- Central database for research, analysis, research and reports,

- Possibility to configure the system with a central database and access to them from network clients,

- Possibility to configure the system as a single workstation as well as with a license server and floating licenses,

- Possibility to install an unlimited number of network clients,

- The function of pre-processing cardiac examinations consisting in automatic recognition and outlining of the contours of the left and right ventricles of the heart based on machine learning algorithms in the CMR analysis module,

- Optimized operation, the ability to run several specialized modules at the same time,

- Possibility of archiving results, reports, strokes / contours of the heart to the PACS system,

- Ability to import previously made strokes / contours of the heart,

- Reporting results, with the possibility of configuring the report,

- Ability to save the report in PDF format and as a text file, as well as XML and JSON,

- Ability to save DICOM SR for right and left ventricular functional measurements,

- Ability to create user accounts to log in to your own profiles,

- Option to export videos in .MPEG or .AVI format,

- Automatic loading of angiographic series into the application,

- Automatic calibration based on isocenter calibration data from the DICOM header,

- Switching the application to full screen mode during arterial analysis,

- 2D and 3D viewing of arteries and lesions,

- The full analysis process is visible at all its stages,

- Application module for determining the optimal angles of acquisition of the second scene necessary for proper 3D modeling during the procedure,

- Automatic search for frames in the ED phase using AI both in scenes synchronized with the ECG curve and without synchronization,

- Automatic preselection module of angiographic series showing only series differing from each other by at least 25 degrees,

- Automatic calculation of optimal viewing angles,

- Automatic segmentation of arterial contours in 2D image based on proven 2D QCA algorithms,

- Automatic 3D reconstruction of coronary artery contours,

- Automatic 2D and 3D reconstruction of surfaces and reference images,

- Automatic quantitative assessment of changes in 3D reconstruction (3D QCA),

- Calculation of shortening of the lesion in the artery for the original 2D views and the current 3D view,

- The system automatically displays the results of the 3 most important lesions,

- The system calculates the following parameters for the lumen of the vessel and atherosclerotic plaque: Diameter and area of the stenosis, Minimum diameter of the lumen of the vessel (MLD), Minimum and maximum diameter for markers P and D (Proximal marker - P and distal - D), Display of reference volume in 3D reconstruction along the entire segment, Narrowing length, Bending angle, Five optimal viewing angles with minimal shortening of the lesion,

- Generating a "pull-back" QFR graph along the artery...... due to restrictions, more information can be found in the SWZ.

PACKAGE NO. 14 – Device for closing access points on the femoral artery

Lp Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Device for closing access sites on the femoral artery - pcs. 30

2 Depth Gauge - Op. 6

1 pcs - 5pcs

TOGETHER:

1.Femoral artery access closure device: 14F – after using devices or shirts in size 10-14F; 18F – after using devices or T-shirts in size 15-18F. Haemostasis achievable by means of a collagen-polymer "sandwich" dressing and additional stimulation of the coagulation process due to the presence of collagen. The closing lock should contain an absorbable collagen dressing, a stainless steel capping element and a spigot of absorbable polymer.

All elements of the closing lock connected on a permanently non-absorbable seam. Polymer absorption time 6 months,

2. Depth meter - a tool to measure the distance between the hole in the artery and the surface of the skin

- available in size 14Fr

- 14 Fr meter catheter made of LDPE and HDPE material.

PACKAGE NO. 15 – Balloon catheters for intra-aortic counterpulsation

Lp Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Balloon catheters for intraaortic counterpulsation - Pcs. 20

TOGETHER:

Intra-aortic contraxation balloon catheters compatible with Arrow and Datascope/Getinge units with the following parameters:

- Balloon catheters in volumes of 30 ml and 40 ml with a profile of 7.5 F with the possibility of seldinger insertion and surgically (with or without a T-shirt),

- Special, coaxial/coaxial catheter structure ensures minimization of resistance to helium flow,

- The central channel reinforced with a spiral braid makes it very resistant to kinks and bends in tortuous arterial vessels, facilitating insertion. In addition, the entire balloon is visible in X-rays,

- Balloon catheter equipped with a hemostatic shield "peel away", reducing bleeding at insertion,

- Hydrophilic coating of the balloon reducing resistance at introduction,

- Highly damage-resistant patented catheter membrane made of Cardiothane material

PACKAGE NO. 16 – Diagnostic catheter for coronary angiography

L.p. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalogue number

1 Diagnostic catheter for coronary angiography - Pcs. 500

TOGETHER:

- diameters 4,5,or 6Fr length 65,80,90,100 or 110 cm

- two-layer polyurethane wall made of PTFE, inner layer hardened with nylon

- reinforced with double steel braid, traction kmontrola 1:1 high resistance to the collapse of the longitudinal structure

- soft tip, atraumatic for the vessel wall distally 2 cm without reinforcement

- large light with a small outer diameter, inner diameter for 4F-0.041/1.03, for 5Fr -0.047/1.20; for 6Fr -0,051/1,30mm

- contrast flow up to 22 ml/sec, resistance to high contrast pressures up to 1000psi

- with or without side holes

- a large selection of curvatures, including available curvatures for Radial access:

- Tiger I/Tiger II 3,5, 4,0, 4,5, 5,0/BLK 4,0

PACKAGE NO. 17 – Drug-coated balloon catheter (paclitaxel)

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Drug-coated balloon catheter (paclitaxel) - pcs. 500

TOGETHER:

- Paclitaxel-coated balloon catheter without polymer carrier,

- Balloon shell in 360° technology;

- Homogeneous coverage: 3-D Dextran-Paclitaxel hydrogel matrix,

- Active substance: Paclitaxel min 2.2μg/mm2,

- Lengths (7): 10-13-15-20-25-30-40 mm,

- diameters (8): 1.5-2.00-2.25-2.50-2,75-3.00-3.50-4.00 mm,

- flexible balloon catheter, low profile made in Vortex technology,

- NP pressure: 6atm/bar,

- Ciśnienie RBP: 14atm/bar,

- Flexible balloon tip (tip) 0.017,

- Shaft proksymalny: 2,2F,

- Shaft dystalny: 2,6F,

- Useful length of the catheter min 135cm,

- Balloon profile for Ø3mm <1mm,

- Balloon material: Nylon 3-fold,

- Semi-susceptible balloon catheter type RX,

- Guide catheter 5F, 0.014" guide

- TDM (Target Deposition Modus) - the drug is released only at higher pressure and contact with the vessel wall,

- Long Lasting EFFECT - the concentration of the drug in the vessel wall persists for several weeks (intima, media, adventitia); up to 60 days

PACKAGE NO. 18 – Balloon catheter not susceptible

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Invulnerable balloon catheter - 200 pcs

TOGETHER:

Balloon catheter invulner not susceptible

• Useful length of the catheter 140 cm,

• Compatible guide catheter: 5F,

• Balloon tip input profile 0.016",

• Balloon folding method - 3-fold balloon,

• Balloon material-Nylon,

• System: Rapid Exchange (RX);

• Hydrophilic balloon shell,

• NP pressure: 12 atm,

• RBP pressure: 22 atm (2.00 to 4.00) and 20 atm (4.50 to 5.00),

• ABP pressure: 30 atm,

• Diameters: 2.00, 2.25, 2.50, 2.75, 3.00, 3.25, 3.50, 3.75, 4.00, 4.50, 5.00 mm,

• Lengths: 6, 8, 10, 12, 15, 18, 20, 22, 25, 30 mm,

• Crossing profile dla 3mm: 0,025”,

• Szaft proksymalny 2.0 F,

• Szaft dystalny: 2.7 F,

• tip length 2.0, 2.5 mm,

• Type of markers - 2 platinum-iridium markers,

• Advanced hydrophilic coating resistant to scratches and damage,

• The material from which the balloon is made allows for high pressure expansion while controlling the diameter and length of the balloon,

• Minimal axial growth of the balloon to eliminate stent distortion,

• High-pressure, minimally susceptibility balloon expansion catheter for coronary angioplasty procedures in a rapid system

Rx exchange

PACKAGE NO. 19 – Cutting balloon catheter

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Cutting balloon catheter - pcs. 30

TOGETHER:

• Nominal pressure - 6 atm.

• Ciśnienie RBP - 12 atm.

• Z-Glide hydrophilic coating

• Crossing Profile 0.036 in (0.91 mm)

• Number of atherosomes:

o 3 for balloons with a diameter of 2.0 – 3.25 mm

o 4 for balloons with a diameter of 3.5 – 4.0 mm

• Lesion entry profile diameter is 0.017" (0.43 mm)

• Catheter working length: 142 -143 cm

• Blade height is 0.005"

• Compatible with 0.014" chain

• Profile szaftu proksymalnego 1.8F

• Profile szaftu dystalnego

o diameter 2.0 – 3.25 mm 2.4F

with a diameter of 3.5 – 4.0 mm 2.7F

• Balloons with diameters from 2.00 mm to 3.25 mm are compatible with 5F guide catheter (0.056" – 1.42 mm)

• Balloons from 3.50 mm to 4.00 mm diameter are compatible with 6F guide catheter (0.066" – 1.68 mm)

Sizes:

Diameters: 2.0, 2.25, 2.5, 2.75, 3.0, 3.25, 3.5, 3.75, 4.0 mm.

Lengths: 6, 10, 15 mm.

Package 2 – Non-compliant balloons

No. Detailed assortment Unit of measurement Quantity per 24 months Net Price Gross Price Net Value Gross Value Manufacturer and catalog number Notes

1 Non-compliant balloons pcs. 500

TOGETHER:

hydrophilic coating, resistant to scratches and damage when tightening the stent;

- system monorail;

- crossing profile 0.027 inch for balloon Ø 3.0 mm

- entry proflie profile 0.018 inch for balloon Ø 3.0 mm;

- nominal pressure min. 12 ATM (for all sizes); -

- RBP pressure 18 atm (for all sizes); - different balloon lengths from 6 - 25 mm;

- different balloon diameters from 1.5 - 5.0 mm changing every 0.25 mm in the diameter range 2.0 - 4.0 mm

PACKAGE NO. 20 – NC cutting balloon catheter

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Cutting balloon catheter - pcs. 20

TOGETHER:

- Non-susceptible (NC) cutting balloon catheter for PTCA procedures,

- The cutting platform is a built-in wire (OD: 0.011") externally integrated and wrapping around the balloon.

- Usable length of the catheter 140 cm,

- Compatible guide catheter: 5F (I.D.>0.056")

- Prowadnik 0,014”

- System: Rapid Exchange (RX);

- Balloon tip input profile: 0.016"

- Hydrophilic balloon shell

- Silicone balloon inner shell

- NP pressure: 12 atm,

- Ciśnienie RBP : 22 atm

- ABP pressure: 30 atm

- Diameters: 2.00, (2.25), 2.50, (2.75), 3.00, (3.25), 3.50, (3,75), 4.00 mm

- Lengths: (6), 8, 10, 12, 15, 18, 20, (23) 25, 30 mm,

- Szaft proksymalny O.D.: 2.0 F,

- Szaft dystalny O.D.: 3.0 F,

- tip length 10mm

- transition profile: 0.035" (3mm), approx. 2.7F

PACKAGE NO. 21 – Cutting and positioning catheter

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Cutting and positioning catheter - pcs. 20

TOGETHER:

Designed for very calcified lesions to adequately increase the lumen of the vessel, prepare the lesion for stenting or use a balloon catheter coated with an antiproliferative drug.

- Equipped with three cutting and positioning elements attached proximal and distal to the balloon

- Balloon design adapted to perform multiple inflations and deflation depending on the length of the shift.

- Szaft proksymalny 2.1F, szaft dystalny 2.7 F

- Nominal pressure 6 atm., RBP 14 atm, balloon length 13 mm, avg. 2.0-4.0mm, minimum 7 sizes.

- Compatible with 0.014" chain, Clinical trial showing TLF of 3% or lower

At least one randomized clinical trial showing significantly lower event rates than conventional POBA

PACKAGE 22 – OTW Microcatheter

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 OTW Microcatheter - pcs 20

TOGETHER:

OTW rotation transfer microcatheter, supporting for 0.014" coronary guides and for injecting precise contrast quantities. 5-layer construction with spiral, oppositely wound double reinforcement, internal supplemented with a PTFE layer along the entire length of the working tip. Available lengths: 135cm and 150cm. Compatible with 5F or larger catheters. Hydrophilic cover 60 cm. Four versions available: Standard (flex), Spiral – with an externally wound plastic spiral on a section of 2cm of the distal part of the catheter, For Hard Changes: with a threaded, metal tip and a spiral, Low profile – with a transition to a 4-layer construction in the distal part. Available diameters: Upholstered tip, 2.6F distal cabinet for Std version, 2.9F distal for Spiral and Hard shifts, 2.2F for low-profile.

PACKAGE NO. 23 – Guide extension cable

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Guide extension cable - pcs. 10

TOGETHER:

Rapid exchange/child in mother extender for backup support, locking 0,014" chain with integral locking balloon. With 0.020" diameter hypotube pusher for 6F, with 13cm hydrophilic coating. Available diameters 6F, 7F, 8F. Equipped with an integral balloon with a marker and rated pressure of 12atm (RBP – 14atm). Tip outer diameter 5.1F for guide catheter 6F. Positioning markers for 95cm and 105cm pusher. Balloon equipped with a golden marker; Spiral reinforced guide extension cable; Usable length 150cm; Internal light 0.056" (for catheter 6F), 0.062" (for catheter 7F), 0.071" (for catheter 8F); Tip outer diameters 0.067" (for 6F catheter), 0.075" (for 7F catheter), 0.085" (for 8F catheter)

PACKAGE NO. 24 – Drug-coated balloon catheter (paclitaxel)

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Drug-coated balloon catheter (paclitaxel) - pcs. 1000

TOGETHER:

- catheter compatible with 0.014" guide, type RX,

- Paclitaxel dose: 3 micrograms per mm2 of balloon,

- carrier of the drug in the form of ammonium salt,

- usable length of the catheter 140cm,

- available balloon diameters: 2,0; 2,25; 2,5; 2,75; 3,0; 3,5; 4 mm,

- available balloon lengths: 15,20,25,30 mm,

- the whole catheter made of polyamide and nylon 12,

- RBP; 16 bar (14 bar for 4.0 x20 and larger),

- effectiveness in small vessels - clinically proven,

- indications for use: restenosis, de novo lesions, changes in small vessels and acute vascular obstruction,

- efficacy confirmed in randomised clinical trials,

- hydrophilic lipophilic coverage in Safepax technology to minimize drug loss during balloon insertion into the vessel and effective implantation of the drug into the vessel

Package No. 3 – Cobalt-chromium coronary stents coated with the latest generation drug with unusual diameters and for difficult changes

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Cobalt-chromium coronary stents coated with the latest generation drug with unusual diameters and for difficult lesions 300 pcs

TOGETHER:

Cobalt-chromium alloy stent based on more than 50% cobalt and iron not exceeding 3% slotted tube

- stent with a solid polymer releasing an analogue of Rapamycin (everolimus)

- stent wall thickness 0.0032"

- crossing profile 0,039”

- length of the supply system 145 cm

- diameter range: 2.0–4.0mm (2.0, 2.25, 2.5, 2.75; 3.0; 3.25, 3.5, 4.0)

- lengths for all diameters: 8, 12, 15, 18, 23, 28, 33, 38 mm

- nominal pressure 9 atm for diameters 2.0-2.5 and 12 atm for diameters 2.75-4.0

- RBP: 16 atm for all sizes

- very good access to sidings

- shortening of the stent after implantation: 0 % (nominal pressure)

- possibility of expanding the stent with a diameter of 3.5 and 4.0 to a diameter of 5.5 mm

PACKAGE NO. 4 – IVUS HD intravascular imaging device with the possibility of performing NIRS spectroscopy

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Bimodal endovascular catheter using two imaging methods simultaneously: Pcs. 400

- ultrasonografię wewnątrznaczyniową – IVUS

- near-infrared spectroscopy - NIRS

• 35-65 MHz extended bandwidth catheter using HD-IVUS technology

• Possibility of simultaneous detection of the structure and composition of atherosclerotic plaque

• Catheter working length: 160 cm

• Imaging range: 150 mm

• IVUS image diameter: 16 mm

• Guide catheter compatible: 6F

• Przejścia profile: 3.2F

2 Device for endovascular imaging using two methods:

- ultrasonografii wewnątrznaczyniowej - IVUS

- near-infrared spectroscopy - NIRS

• Simultaneous recording of IVUS and NIRS imaging data

• Technology for detecting lipid core in atherosclerotic plaques

• Extended IVUS frequency range: 35-65 MHz for sharp HD-IVUS quality

• Rozdzielczość osiowa IVUS: 20 μm

• Prędkość pracy: 0.0, 2.0, 10.0 mm/s, 960 rpm

• Prędkość pullback: 0.5, 1.0, 2.0 mm/s, 1800 rpm

The 24-month lease includes a mandatory annual maintenance inspection of the console.

Due to the personal data processing zone, the servicer will sign a confidentiality statement before performing service activities. M-c 1 pc.

TOGETHER:

Package No. 5 – Seam systems for closing vascular accesses

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Suturing systems for closing vascular accesses Pcs. 60

TOGETHER:

Seam systems for closing vascular accesses under the introducer up to 24F.

A set consisting of 3 elements:

- devices for attaching seams, equipped with a measuring tape of tissue depth;

- knot pusher,

- obcinarki

Possibility to close 5-8F holes with one closing machine and up to 24F with two

Package No. 6 – Treatment chain

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 Treatment guide Pcs. 10

TOGETHER:

Controlable treatment guide diameter 0.035” with upholstered tip 0,025” length 17 cm

Guide length from min. 180cm to min. 300 cm with coating for easy passage through narrowed areas

0.014" and 0.018" chains and a specialist chain for the CTO in the area below the knee are also available.

PACKAGE NO. 7 – Active circulatory pump drawing blood directly from the left ventricle of the heart

No. Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 • Actively supporting circulation pump drawing blood directly from the left ventricle of the heart - pcs. 30

• Capacity: up to 4.3 L per minute

• Cabinet diameter 9F

• Pump diameter 14F

• Compatible with 14F T-shirt

• Compatible with 0.018" chain

• Measurement of total cardiac output

• Pump repositionable without additional imaging

• Optical pressure sensor

• Remote access to pump operation

• Ability to maintain pump support for up to 5 days

2 Lease of pump control units 24 months.

The lease includes a mandatory annual inspection of pump control units

M - c 2 pcs.

TOTAL : (items 1 and 2)

The lessee is obliged to possess, and the lessor to lease two consoles for the duration of the treatments.

This requirement is due to the fact that it is life-saving equipment and it is necessary to ensure its constant availability in the event of a failure of one of the consoles.

The Contractor shall ensure:

- warranty and free service of the device and the leased device;

- free technical inspections with spare parts necessary for the inspection once a year;

- time to remove the failure of devices and leased devices not longer than 5 working days counted from the moment of undertaking the repair;

- in the event of a failure of devices that cannot be removed within 5 working days, the Contractor guarantees the delivery of a replacement device for the time of repair, with parameters not worse than the device that has failed.

PACKAGE NO. 8 – Rail guide for laterals

Lp Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

Chain for platoons Pcs. 20

TOGETHER:

Prowadnik wants collateral

• Diameter 0.014"

• Energy chain made of stainless steel

• Distal part of the chain woven from 19 wires for high resistance and excellent maneuverability and feeling of the chain

• Spiral part with a length of 19 cm

• Guide core made of one piece of wire

• Tip stiffness 0.3g

• Shader tip 3cm (platinum)

• Tip shape: pre-shaped and straight

• Hydrophilic coating on the distal 52cm

• PTFE-coated energy chain cabinet

• Available lengths: 190cm and 300cm

• Guide for complex PCI treatments, especially for going through very tight and winding lesions and for retrograde access (passing through collaterals)

PACKAGE NO. 9 – PTCA chain

Lp Detailed assortment Unit of measurement Quantity per 24 months Net price Gross price Net value Gross value Manufacturer and catalog number Notes

1 PTCA Guide Pcs. 20

TOGETHER:

Prowadnik do PTCA

• Coronary artery chain

• Diameter 0.014 in (0.36 mm)

• Energy chain made of stainless steel

• Distal part of the chain made of 8 wires

• Transfer of turnover 1:1

• Composite core of the chain made of: one piece of wire + braided with 6 wires

• Tip stiffness 3.0 g

• Shader tip 3 cm (platinum)

• Tip shape: bent to a 1 mm section

• Hydrophilic coating on the distal 50 cm

• Polymer jacket on distal 40 cm

• Braid length: 11 cm

• Available lengths: 200 cm, 300 cm

Download full details as .pdf
The Buyer:
Specjalistyczny Szpital im. dra Alfreda Sokołowskiego
CPV Code(s):
33111710 - Angiography supplies