Description:

OUS will procure 2 and have the option to procure 1 (similar to operation 2) single door steam sterilisers (approx. 2 sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.

Sterilisation device/container dimensions:

(L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15 mm),(L60 mm x B 40 mm x H20 mm),(L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34,5 cm.



OUS will procure 2 and have the option to procure 1 (similar to operation 2) single door steam sterilisers (approx. 2 sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.

Sterilisation device/container dimensions:

(L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15 mm), (L60 mm x B40 mm x H20 mm), (L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34,5 cm.

External measurement are preferably the same for both of the surgical departments but is a challenge due to the container size.

Operating theatre 1 D2.4052:

Maximum space is:

Depth: enter the outer measurement depth (must be able to take container L70 mm x B28 mm x H20 mm, L60 mm x B28 mm x H13 mm), (L54B x B26 mm x 15 mm)

Width: 900 mm — 1 000 mm

Height: can be a little higher than 1 970 mm

Operating theatre 2 D3.4023:

Maximum space is:

External measurements:

Depth: 1 020 mm (must be able to sterilise the sterilisation device (must be able to hold the sterilisation device: (L58 mm x B28 mm x H21 mm), (L62 mm x B42 mm x H20 mm)/(L60 mm x B20 mm x H20 mm)

Width: 900 mm — 950 mm

Height: can be a little higher than 1 970.

The steam steriliser delivered must be able to be transported into the named sterilisation room without building intervention (structural works). OUS will not perform any alteration of doors/transport routes in the corridors to get the sterilisers into the sterilisation room, due to patient activity.

Transport route:

Door into the sterilisation room

Operating theatre 2: 87.5 cm x 203.5 cm

Operating theatre 1: 98 cm x 209 cm

One extra steam steriliser at Rikshospitalet [the National Hospital] needs to be replaced and we would like a price for 1 device as an option.