Laboratory, optical and precision equipments (excl. glasses) | Tenderlake

Laboratory, optical and precision equipments (excl. glasses)

Contract Value:
GBP 150K -
Notice Type:
Contract Notice
Published Date:
12 October 2022
Closing Date:
11 November 2022
Location(s):
UK UNITED KINGDOM
UKH EAST OF ENGLAND (UK UNITED KINGDOM)
Description:
CEFAS22-53 Invitation to Tender for provision of a chemical imaging and spectral analysis system for fast screening of microplastics in environmental samples down to at least 20 m, with requirements including small footprint, Quantum Cascade Laser (QCL) technology, automation, no cooling liquids, ability to image large samples and operate in reflectance and attenuated total reflectance (ATR) mode, support samples on filters, analysis of particles down to at least 20 m, software with easy method creation and export options, and delivery before or during WC 20 March 2023.
CEFAS22-53 Invitation to Tender for provision of a chemical imaging and spectral analysis system

The Marine Litter Team at Cefas require a chemical imaging and spectral analysis system to maintain and enhance Cefas’ capability for the fast screening of microplastics in environmental samples down to at least 20 m. The proposed system will include an integrated particle characterisation process including size of the particles as well as polymer type by comparison to a built-in library database.

The Marine Litter Team at Cefas require a chemical imaging and spectral analysis system to maintain and enhance Cefas’ capability for the fast screening of microplastics in environmental samples down to at least 20 m. The proposed system will include an integrated particle characterisation process including size of the particles as well as polymer type by comparison to a built-in library database.

Chemical Imaging and spectral analysis system requirements:

• have a small footprint as laboratory space is limited. The unit should have a maximum footprint of (W x H x D) of 1000 x 500 x 650 (mm)

• be based on Quantum Cascade Laser (QCL) combined with scanning optics

• be highly automated

• not require cooling liquids such as liquid nitrogen, to reduce operating costs

• able to image large samples (e.g., 25 mm filters) as well as focusing on specific zones/particles of interest

• be able to operate in reflectance and attenuated total reflectance (ATR) mode and should be able to automatically switch between the two modes

• support samples on filters applications (e.g., gold, silver or other types of filters) of at least 25 mm diameters

• be able to analyse particles down to at least 20 m in size

• be provided with a PC with Windows 10 as a minimum and operating software including a full polymer library

• include all other related key components such as nitrogen or air compressors

Software requirements:

• offer software with a fast and easy method creation

• include additional features including spectra post-processing options

• include search libraries for reliable compound identification with the option of creating in-house libraries

• allow for easy exports of data (e.g., MS Excel format) including analytical spectra and library matches

It is important to Cefas that delivery is made before or during WC 20 March 2023 to ensure payment by end of FY22-23. If delivery cannot be accommodated in this timeframe, please refrain from bidding.

Download full details as .pdf
The Buyer:
Cefas
CPV Code(s):
38000000 - Laboratory, optical and precision equipments (excl. glasses)