• CE certified system that can still be delivered in 2024.
• Despite compact design (installation on surface for standard PET/CT systems possible)
Large-field PET detector greater than 100 cm depth for simultaneous display of the entire
trunk incl. head in a bed position.
• Highest effective sensitivity of the PET system of more than 900 cps/kBq for shortest
Examination times of less than one minute and/or examinations with the lowest
Activity levels and thus radiation exposures of patients and staff.
• Best spatial resolution of less than 4 mm (reconstructed with filtered rear projection) for
Visualization of the smallest lesions and structures
• High-performance CT component for dynamic CT examinations with and without
Contrast media including CT-Dual Energy examinations in a spiral and use CT data
with an arbitrary kV value for the reconstruction of images that the
values of relative electron density and/or mass density in order to
Use radiotherapy planning
• Provision of a flat table support for planning for external irradiation in
Connection with a gantry opening of more than 75 cm and a laser positioning system for
Marking of the lsocenter
• Full compatibility with the existing network infrastructure at UKM and the
existing post-processing and evaluation platform Syngo.Via
• LSO crystal-based system with measured time-of-flight with high temporal resolution < 250 ps. As a result, high spatial resolution, significant improvement of reconstruction and reduced dispersion in obese patients (important for clinical and methodological and clinical research projects).
Please note: This announcement is purely an expression of interest procedure (market exploration) and not the start of an award procedure aimed at the award of a contract!
LOT-0001
Supply and installation of a PET/CT with long axial field of view.
A system with a long axial field of view is to be procured in the intended configuration (required
Length of axial field of view, detector material, configuration and sensitivity, as well as
Recording mode and software configuration) with the following features:
• CE certified system that can still be delivered in 2024.
• Despite compact design (installation on surface for standard PET/CT systems possible)
Large-field PET detector greater than 100 cm depth for simultaneous display of the entire
trunk incl. head in a bed position.
• Highest effective sensitivity of the PET system of more than 900 cps/kBq for shortest
Examination times of less than one minute and/or examinations with the lowest
Activity levels and thus radiation exposures of patients and staff.
• Best spatial resolution of less than 4 mm (reconstructed with filtered rear projection) for
Visualization of the smallest lesions and structures
• High-performance CT component for dynamic CT examinations with and without
Contrast media including CT-Dual Energy examinations in a spiral and use CT data
with an arbitrary kV value for the reconstruction of images that the
values of relative electron density and/or mass density in order to
Use radiotherapy planning
• Provision of a flat table support for planning for external irradiation in
Connection with a gantry opening of more than 75 cm and a laser positioning system for
Marking of the lsocenter
• Full compatibility with the existing network infrastructure at UKM and the
existing post-processing and evaluation platform Syngo.Via
• LSO crystal-based system with measured time-of-flight with high temporal resolution < 250 ps. As a result, high spatial resolution, significant improvement of reconstruction and reduced dispersion in obese patients (important for clinical and methodological and clinical research projects).