Successive delivery of single-cavity stimulators, unicavity cardioverters, dujam cardioverters, cardioverters with resynchronization function

Successive deliveries of single-cavity pacemakers, single-cavity cardioverters, dujame cardioverters, cardioverters with resynchronization function, electrode insertion kits, electrodes for temporary cardiac stimulation

Single-cavity stimulators with repair kit

1. "Single-cavity stimulator, year of manufacture not earlier than 2022, weight max. 30g, pulse amplitude min range 0.5 - 6.0V, pulse width (A/V) min range 0.5 - 1.0 ms, ventricular sensitivity at least in the range of 1.0 - 7.5 mV, vestibular sensitivity at least in the range of 0.25 - 4.0 mV, passive and active A and V electrodes to choose from, steroid A and V electrodes, night program, A/V refractive period min range 250 - 400 ms, recording electrode resistance trends throughout the life of the device, function automatically determining the ventricular stimulation threshold and automatically adjusting the ventricular stimulation parameters to the measured stimulation threshold, the possibility of programming V refraction above 400 ms, the possibility of programming V refraction below 200 ms, MRI system (pacemaker with electrodes) approved for testing 1.5 and 3.0 tesla without exclusion zones. Deposit 5 pcs.

Note: The Ordering Party requires that the Contractor, for the duration of the contract, provides the Ordering Party with a device for measuring intraoperative stimulation parameters together with two sets of cables and pacemaker control after implantation free of charge. Note: The ordering party requires the provision of passports and patient information/instructions

"80 pieces

2. "Endocavitary electrodes for stimulation, A – vestibular V – ventricular electrodes, passive and active to choose from, steroid electrodes A and V.

Deposit 5 pcs.

"80 pieces

3. Screwdriver Deposit 3 pcs. pieces 10

4. Mandryn for electrodes - various lengths in the range of 45 cm-65 cm, straight or J shape to choose by the Ordering Party Deposit 4 pcs. pieces 10

5. Key for screwing in the electrode Deposit 4 pcs. pieces 10

6. End cap IS-1, DF-1 port input of the device to be selected by the Ordering Party Deposit 2 pcs. pieces 10

Dual-cavity stimulators with repair kit

1. "Two-cavity stimulator, year of manufacture not earlier than 2022, pacemaker lifetime not less than 6.6 years at 100% stimulation, A = 2.5V & V = 2.5V, pulse width 0.4 ms, stimulation frequency: 60 min-1, minimum amplitude of pulse A and V 0.5 – 7.5V, minimum pulse width A and V 0.5 – 1.0 ms, vestibular sensitivity 0.20 – 4.0 mV, ventricular sensitivity 0.1 – 6.5 mV, minimum atrioventricular interval 30-350 ms, automatic determination of the stimulation threshold and adjustment of stimulation parameters, automatic polarity switching in case of exceeding the electrode impedance range, IEGM recording over 35 sec, passive and active A and V electrodes to choose from, steroid A and V electrodes to choose from, MRI system (stimulator with electrodes) approved for 1.5 and 3.0 tesla tests without exclusion zones.

Deposit 12 pcs.

Note: The Contracting Authority requires that the Contractor, for the duration of the contract, provide the Contracting Authority with a device for measuring intraoperative stimulation parameters together with six sets of cables and pacemaker control after implantation free of charge.

The Contractor undertakes to provide the programmer and paper to the programmer printer free of charge in an amount appropriate to the number of ordered devices. Note: The ordering party requires the provision of passports and information/instructions for patients" pieces 320

2. "Endocavitary electrodes for stimulation, V – ventricle, passive and active electrodes for selection, steroid A and V electrodes.

Deposit 24 pcs.

"640 pieces

3. Screwdriver Deposit 3 pcs. pieces 10

4. Mandryn for electrodes - various lengths in the range of 45 cm-65 cm, straight or J shape to choose by the Ordering Party Deposit 4 pcs. pieces 10

5. Key for screwing in the electrode Deposit 4 pcs. pieces 10

6. End cap IS-1, DF-1 port input devices to be selected by the Ordering Party 10 pieces

Single-cavity cardioverters with repair kit

1. "Cardioverter/single-cavity defibrillator, year of manufacture not earlier than 2022, DFI standard, DF4 device life at 15% pacing, 2 or more charges to max energy per year, 500 Ohm, 2.5V over 6.0 years, sterilization min 6 months, defibrillation energy delivered min 35J, number of shocks in one intervention min 6, automatic description of battery status and electrode resistance, pulse amplitude min. range 0.5 - 5.0 V, pulse width (A/V), min. range 0.5 - 1.0 ms, chamber sensitivity - at least in the range of 0.2 - 1.2 mV, recording electrode resistance trends, ATP in VF, possibility of excluding the can from the HV cycle, possibility of excluding SVC COIL from the HV cycle.

Deposit 5 pcs. REMARK!!! The Contracting Authority requires that the Contractor for the duration of the contract provides the Ordering Party with the following devices: a device for measuring intraoperative stimulation parameters (together with two sets of cables) and cardioverter control after implantation. Note: The ordering party requires the provision of passports and patient information/instructions

"100 pieces

2. "Endocavitary electrodes for defibrillation, steroids, passive/active defibrillation electrodes – to choose from, single-koyl and two-koil electrodes – to choose from,

Deposit 5 pcs.

"100 pieces

3. Screwdriver Deposit 3 pcs. pieces 10

4. Mandryn for electrodes - various lengths in the range of 45 cm-65 cm, straight or J shape to choose by the Ordering Party Deposit 4 pcs. pieces 10

5. Key for screwing in the electrode Deposit 4 pcs. pieces 10

6. End cap IS-1, DF-1 port input of the device to be selected by the Ordering Party Deposit 2 pcs. pieces 10

Cardioverter two-cavity with repair kit

1. "Cardioverter/two-cavity defibrillator, year of manufacture not earlier than 2022, device life at 15% pacing, min 2 discharges or more charges to max energy in a year, 500 Ohm, 2.5V over 6.0 years, sterilization min 6 months, defibrillation energy delivered min 35J from the first therapy, number of shocks in one intervention min 6, automatic description of battery status and electrode resistance, pulse amplitude min 0.5 - 5.0 V, pulse width (A/V), min range 0.5 - 1.0 ms, ventricular sensitivity - at least in the range of 0.2 - 1.2 mV, vestibular sensitivity - at least in the range of 0.2 - 4.0 mV, ATP in VF, possibility of excluding SVC coil from the HV circulation, possibility of excluding the can from circulation,

Deposit 3 pcs.

REMARK!!! The Contracting Authority requires that the Contractor for the duration of the contract provides the Contracting Authority with the following devices: a device for measuring intraoperative stimulation parameters (with two sets of cables) and cardioverter control after implantation Note: The Ordering Party requires the provision of passports and information/instructions for patients

"20 pieces

2. "Endocavitary electrodes for stimulation, A – vestibular electrodes, passive and active to choose from, steroid electrodes A and V.

Deposit 3 pcs.

"20 pieces

"Endocavitary electrodes for defibrillation, steroids, passive/active defibrillation electrodes – to choose from, single-koyl and two-koyl electrodes – to choose from

Deposit 3 pcs.

"20 pieces

3. Screwdriver Deposit 3 pcs. pieces 10

4. Mandryn for electrodes - various lengths in the range of 45 cm-65 cm, straight or J shape to choose by the Ordering Party Deposit 4 pcs. pieces 10

5. Key for screwing in the electrode Deposit 4 pcs. pieces 10

6. End cap IS-1, DF-1 port input of the device to be selected by the Ordering Party Deposit 2 pcs. pieces 10

Cardioverters – CRT-D resynchronisation cardiac defibrillators with equipment

1."Cardioverters – cardiac defibrillators with CRT-D resynchronization function:Weight ? 85 grams, Delivered defibrillation energy ?35J, Two-phase pulse, Minimum 2 shock vector configurations, ATP therapy in the VF zone, Number of high-energy discharges in one sequence for each zone (min. 5), Possibility to program sensitivity to avoid sensing ... T, VV delay (0-80 ms), Wireless communication with programmer, Alternatively delivery of the device with DF-1/DF-4;IS-1/IS-4 socket, Wireless ECG, Automatic follow up including automatic stimulation threshold measurement, measurement of own potentials, measurement of battery status impedance, Wireless communication with the programmer, Telemonitoring stations with data transmission in the GSM system, Minimum 3 left ventricular stimulation configurations, Possibility of rata response stimulation during the operation of the Mode Switching algorithm, Configurable IEGM channel recording Deposit 3 pcs. The manufacturer provides 8 sets of cables to the analyzer free of charge

REMARK!!! The Contractor undertakes to provide free of charge the programmer and paper to the programmer printer in an amount appropriate to the number of ordered devicesNote: The Ordering Party requires the provision of passports and information / instructions for patients" pieces 30 2. "Electrodes for defibrillation active, single-koil/two-koyl – to choose from, all steroid electrodes Electrodes with DF-1/DF-4 connector to be chosen by the Customer Electrode diameter < 9F Deposit 3 pcs." pcs 30

3. "Vestibular electrodes: Steroid electrodes with IS-1 connector, Shape: straight, Passive/active electrodes to choose from, length 50 cm – 60 cm (at the choice of the Ordering Party) Deposit 3 pcs." pieces 30 4. LV ventricular stimulation electrode (for coronary sinus):Bipolar electrode with IS 1 connector Electrode with silicone or polyurethane sheath Four-pole electrode "over the wire" with IS-4 connector Deposit 3 pcs. pieces 30 5. Guide for left chamber electrode Angioplastic type guide At least 2 types of chain hardness Straight and curved chain end (J) pieces 20

6. Set for insertion of the electrode into the coronary sinus: Different curvatures of "shirts" into the coronary sinus "Integrated valve" for the above-mentioned shirts Balloon catheter for coronary sinus venography 40 pieces

7. 4-pole diagnostic catheter: Catheter diameter max. 6F, Useful length min. 110 cm Available curvatures (minimum) : Josephson, Cournand Electrode rings 1-2 mm wide, Ring spacing 2-5-2 mm, Electrode rings made of platinum pieces 20

8. Intorducer 8F/9F to choose from: sheath, delator and guidewire 40 pieces

9. Screwdriver pieces 6

10. Guide of pieces 6

11. Hood pieces 6

12. End cap 6 pieces

Puncture and electrode insertion kit

A set for puncturing and inserting stimulation or endocavitary defibrillation electrodes into the right heart cavity, consisting of a sleeve, dilator, guide, needle and a 10 ml three-piece syringe. Luer-Lock, in compatible sizes, year of manufacture not earlier than 2022, tearable vascular sleeve offered in sizes from 7F to 11F (to be chosen by the Ordering Party). pieces 400

External electrode - endocavitary for temporary stimulation of the heart

"External electrode - endocavitary for temporary cardiac stimulation, size 6F, distal end type: soft, bent, sterile, non-toxic, apirogenic.

"30 pieces