Minimum requirements for the instrument:
• ionization source is capable of operating both in positive/negative ion modes;
• possibility to operate in linear and reflectron mode;
• pulsed ion extraction;
• mass range extending at least up to 500 kDa;
• mass resolution of 1000 FWHM or higher in linear mode and 20 000 FWHM or higher in reflectron mode;
• mass accuracy of 200 ppm or better in linear mode and 50 ppm or better in reflectron mode using external calibration;
• MS-MS capability;
• software for advanced polymer analysis and basic protein analysis, 2 licenses;
• the system must have warranty at least for 12 months;
• the policy for software upgrades and their pricing must be included;
• the instrument must be new one. No prototypes allowed;
• 10 steel target plates;
• a windows 10 computer to operate the instrument with all the necessary programs installed;
• the instrument is CE-marked.